Skip to main content

Research Repository

Advanced Search

Assessment of a non-invasive haemoglobin sensor NBM 200 among pregnant women in rural India

Ahankari, A S; Fogarty, A W; Tata, L J; Dixit, J V; Myles, P R

Authors

A S Ahankari

A W Fogarty

L J Tata

J V Dixit

P R Myles



Abstract

© 2016, BMJ Publishing Group. All rights reserved. Objective This study aimed to assess a noninvasive haemoglobin (Hb) sensor NBM 200 in pregnant women in a rural Indian setting. Methods The study population consisted of women between 3 and 5 months of pregnancy, from 33 villages in Tuljapur and Lohara blocks of Osmanabad district, Maharashtra, between April 2014 and June 2015. Hb measurements obtained from the non-invasive sensor NBM 200 were compared with measurements obtained from an automated haematology analyser Sysmex XP-100, using the Bland Altman method and Spearman’s Rank correlation coefficient. Interclass correlation coefficient (ICC), sensitivity and specificity values were used to assess the anaemia diagnostic accuracy of NBM 200 against the gold standard (Sysmex XP-100). Results Data were obtained from 269 pregnant women (median age: 21 years, IQR: 19– 23 years). Hb levels estimated by the Sysmex XP- 100 analyser ranged from 5.5 g/dL to 14.1 g/dL (mean: 10.0 g/dL, SD: 1.28), while measurements obtained from NBM 200 ranged from 9.5 g/dL to 14.6 g/dL (mean: 11.9 g/dL, SD: 1.43). Spearman’s test found a significant, moderately positive correlation between the two methods (r s =0.4, p < 0.001), ICC was 0.22, and the Bland-Altman analysis showed a mean difference of −1.8 g/dL (95% CI −2.06 to −1.71), indicating a systematic overestimation of Hb using the NBM 200. The NBM 200 showed low sensitivity (33.7%; 95% CI 27.3% to 40.5%) but high specificity (91.8%; 95% CI 81.9% to 97.3%) for the diagnosis of anaemia. Conclusions Hb measurements obtained from the NBM 200 were higher with consequent underestimation of anaemia as compared with the gold standard reference method. This limits the use of the NBM 200 as an anaemia diagnostic test in our study population consisting of women during pregnancy.

Citation

Ahankari, A. S., Fogarty, A. W., Tata, L. J., Dixit, J. V., & Myles, P. R. (2016). Assessment of a non-invasive haemoglobin sensor NBM 200 among pregnant women in rural India. BMJ Innovations, 2(2), 70-77. https://doi.org/10.1136/bmjinnov-2015-000085

Acceptance Date Jan 16, 2016
Online Publication Date Feb 3, 2016
Publication Date 2016-04
Deposit Date Feb 22, 2018
Publicly Available Date Mar 29, 2018
Journal BMJ Innovations
Electronic ISSN 2055-642X
Publisher BMJ Publishing Group
Peer Reviewed Peer Reviewed
Volume 2
Issue 2
Pages 70-77
DOI https://doi.org/10.1136/bmjinnov-2015-000085
Keywords General Medicine
Public URL https://hull-repository.worktribe.com/output/596886

Files





You might also like



Downloadable Citations