S. Nichols
Variability of automated carotid intima-media thickness measurements by novice operators
Nichols, S.; Milner, M.; Meijer, R.; Carroll, S.; Ingle, L.
Abstract
Carotid intima-media thickness (C-IMT) measurements provide a non-invasive assessment of sub-clinical atherosclerosis. The aim of the study was to assess the inter- and intra-observer variability of automated C-IMT measurements undertaken by two novice operators using the Panasonic CardioHealth Station. Participants were free from cardio-metabolic disease and each underwent serial bilateral C-IMT ultrasound measurements. Immediate inter operator measurement variability was calculated by comparing initial measurements taken by two operators. Immediate retest variability was calculated from two consecutive measurements and longer-term variability was assessed by conducting a further scan one week later. 50 apparently healthy participants (n=20 females), aged 26.2 ± 5.0 years were recruited. Operator 1 recorded a median (inter-quartile range) right and left-sided C-IMT of 0.471mm (0.072mm) and 0.462mm (0.047mm). Female’s right and left C-IMT was 0.442mm (0.049mm) and 0.451mm (0.063mm) respectively. The limits of agreement (LoA) for immediate inter-operator variability were -0.063 to 0.056mm (mean bias -0.003mm). Operator 1’s immediate retest intra-operator LoA were -0.057 to 0.046mm (mean bias was -0.005mm). One week LoA were -0.057 to 0.050mm (mean bias -0.003mm).Operator 2 recorded median right and left sided C-IMT of 0.467mm (0.089mm) and 0.458mm (0.046mm) for males respectively whilst female measurements were 0.441mm (0.052mm) and 0.444mm (0.054mm) respectively. Operator 2’s intra-operator immediate retest LoA were -0.056 to 0.056 (mean bias
Citation
Nichols, S., Milner, M., Meijer, R., Carroll, S., & Ingle, L. (2016). Variability of automated carotid intima-media thickness measurements by novice operators. Clinical physiology and functional imaging, 36(1), 25-32. https://doi.org/10.1111/cpf.12189
Journal Article Type | Article |
---|---|
Acceptance Date | Aug 15, 2014 |
Online Publication Date | Sep 12, 2014 |
Publication Date | 2016-01 |
Deposit Date | May 13, 2015 |
Publicly Available Date | May 13, 2015 |
Journal | Clinical physiology and functional imaging |
Print ISSN | 1475-0961 |
Publisher | Wiley |
Peer Reviewed | Peer Reviewed |
Volume | 36 |
Issue | 1 |
Pages | 25-32 |
DOI | https://doi.org/10.1111/cpf.12189 |
Keywords | Reliability, Reproducibility, C-IMT, Ultrasound |
Public URL | https://hull-repository.worktribe.com/output/373710 |
Publisher URL | http://onlinelibrary.wiley.com/doi/10.1111/cpf.12189/abstract |
Additional Information | Author's accepted manuscript of article published in : Clinical physiology and functional imaging, 2014 at http://onlinelibrary.wiley.com/doi/10.1111/cpf.12189/abstract |
Contract Date | May 13, 2015 |
Files
Article.pdf
(596 Kb)
PDF
Copyright Statement
© 2015 University of Hull
You might also like
A clinician's guide to cardiopulmonary exercise testing 1: An introduction
(2015)
Journal Article
A clinician's guide to cardiopulmonary exercise testing 2: test interpretation
(2015)
Journal Article
Downloadable Citations
About Repository@Hull
Administrator e-mail: repository@hull.ac.uk
This application uses the following open-source libraries:
SheetJS Community Edition
Apache License Version 2.0 (http://www.apache.org/licenses/)
PDF.js
Apache License Version 2.0 (http://www.apache.org/licenses/)
Font Awesome
SIL OFL 1.1 (http://scripts.sil.org/OFL)
MIT License (http://opensource.org/licenses/mit-license.html)
CC BY 3.0 ( http://creativecommons.org/licenses/by/3.0/)
Powered by Worktribe © 2024
Advanced Search